Data logger

ESS3 R1

Measuring device with radial pressure sensor

The ESS3 R1 series devices measure pressure, differential pressure, and temperature in gas, water, and district heating networks and are equally suitable for liquid and gaseous media in industrial applications.

The battery-operated, modular devices consist of a control unit (housing, processor, software, display), sensor, and battery.

They are approved for hazardous areas (Zones 1 and 2) and are waterproof up to protection class IP68.

Measurement parameters

Typical applications

The ESS3 MODEM2024 is particularly suitable for:

Fault analysis in the gas network

Monitoring in gas pressure regulating and measuring systems

Pipeline network calculation

Leak testing G469 / W400-2

Sensor

The sensor is the metrological link to the application. Performance and user-friendliness are therefore the main focus:

Sensor replacement by the user is possible and immediately ready for operation without recalibration

Stainless steel encapsulated, piezoresistive sensor with high long-term stability, resistant to aggressive media

High resolution of measured values; multiple measuring ranges possible for a single sensor

Measurement of media temperature

High measuring rates due to high natural resonance frequency

High overpressure safety and high burst pressure

Special versions, e.g., for O2 measurement

Sensibly graduated fixed as well as user-definable measuring ranges and various accuracy classes up to +- 0.05% of FSD (full scale deflection)

Temperature sensors suitable as rod probes or for use in thermowells

Technical Data

ApplicationMeasurement and Storage of Data (Pressure and Temperature) for Fault Analysis, Monitoring of Pipeline Networks and Gas Pressure Control Systems
Sensor ConnectionsOne radial sensor connection (M30) for connecting a pressure or temperature sensor
Explosion Protection ClassEx II 2G Ex ib IIC T4 Gb
Protection Classes, HousingDepending on the sensor: IP67 for gauge pressure, IP68 for absolute and differential pressure as well as temperature; W × H × D [mm]: 108 × 162 × 80; Weight [kg]: 1.2
Pressure Sensor Measurement RangesGauge Pressure: 0 … 100/250 mbar and 0 … 1/2.5/10/25/100 bar; Differential Pressure: 0 … 100 mbar, 0 … 1/10 bar; Additional measuring ranges available on request
Temperature Sensor Measurement Ranges-10 °C … +40 °C and -30 °C … +150 °C
Measurement Interval125 ms … 6 h
Initial Measurement AccuracyDepending on the sensor (up to 0.05% FS)
ResolutionUp to 0.004% FS
Communication InterfacesIrDA; Display; Keypad
Operating DataBattery operation up to 8 years
DisplayCurrent Value; Maximum and Minimum Value as well as Differential Value; Memory Usage and Battery Status
SettingsTime and Date; Upper and Lower Alarm Threshold; Averaging (2 … 600 values); Resolution; Measurement Location Name (29 characters); Storage Method (rolling/static)
OperationVia menu using keypad; Via TfsWin III software using IrDA interface cable
Data Storage250,000 date-time values / 512 kB
Typical Range2 years (through data compression)
SoftwareTfsWin III for parameterization, visualization, analysis and archiving of data

Service

We will answer your questions in the administrative and commercial area.

+49 721 680381-30

Support

For technical support for your product, you can open a support ticket.

Sales

The sales department is your central contact for product inquiries and quotation preparation.

+49 721 680381-20

Downloads

Here you will find various downloads to support you.

Brochures, data sheets and application reports – we have all the relevant information for you here.

Blog

News from the UNION world.

Lots of news from the world of measurement technology.

Products

Special products for your industrial application.

Our range includes CWD calorimeter devices, INCA analyzers, ESS devices and much more.

Scroll to Top